STABILITY CHAMBERS
Installation, Operation and
Maintenance Instructions
GENERAL
2
Inspection 2
Location 2
Installation 2
Door Alignment
2
Shelf Installation
2
Remote Contacts
2
RS485 Port
2
Internal Outlet
2
Access Port
2
Humidity Installation 3
Drain Installation 3
Electrical 4
Glycerin Well Assembly 4
OPERATION
4
MAINTENANCE
5
Periodic Cleaning 5
Cleaning the Humidifier 6-7
Maintenance Service and
Analysis Guide
8
1/10 Rev. E 132574
Seitenansicht 0
1 2 3 4 5 6 7 8

Inhaltsverzeichnis

Seite 1 - Maintenance Instructions

STABILITY CHAMBERS Installation, Operation and Maintenance Instructions GENERAL 2 Inspection 2 Location 2 Installation 2 Door Alignment 2

Seite 2 - GENERAL

GENERAL INSPECTION When the equipment is received, all items should be carefully checked against the bill of lading to insure all crates and cartons

Seite 3 - - 3

Duplex/ or European Outlet - (Optional) is located near the top of the interior cabinet back. This is a 15 amp 115 volt duplex PN 88010300 with its o

Seite 4 - OPERATION

An 18” length of 1/2" I.D. drain hose is shipped inside cabinet. Attach hose to drain stub under cabinet with supplied clamp and run to an atmo

Seite 5 - MAINTENANCE

- 5 - The cabinets utilize electrically operated heaters to warm the cabinets in the heating mode. The heating elements are located in the interi

Seite 6 - Figure B

CLEANING THE HUMIDIFIER Before Cleaning: 1. Disconnect the electrical plug from power source. Turn off water supply. (See Figure D) 2. Unfasten th

Seite 7 - - 7

- 7 - 4. Clean the Atomizing Unit. Do not submerge in water. a. Slightly twist the Cylindrical Screen out of the lock position and remove. b.

Seite 8 - - 8

- 8 - MALFUNCTION POSSIBLE CAUSE SOLUTION Compressor will not start - 1. Service cord unplugged 1. Plug in service cord no hum 2. Fuse bl